For Voltage inputs
Uses 0-100 mV input range resolved in
12 bit.
For higher input ranges, use externally calibrated
shunt or contact factory
Total error: ± 0.1% of scale ± 1 bit.
For Current inputs
Uses 0-20 mA input range with 5 ohms
precision shunt resolved in 12 bit.
Total error: ± 0.1%
of scale ± 1 bit.
For RTD input
Uses 3 wire compensation technique. Current
sense is 0.5mA Max.
Power dissipated
in RTD is 0.025 mV @ 100 ohms.
Range supported: -200 to
850ºC
For Thermocouple input
Uses 0-100 mV input range resolved in
to 12 bit for positive input and resolved
in 11 bit for
negative inputs.
Cold junction error is + 1º maximum
and 0.5º typical.
Total error: ± 0.5% of
scale ± 1 bit + CJC error |